Thoratec Corporation Issues Worldwide Medical Device Correction
of HeartMate II® Left Ventricular Assist System
|
|
Date of Recall:
Friday, October 24th, 2008
Recall Details:
PLEASANTON, CA -- October 24, 2008—Thoratec
Corporation is initiating a worldwide medical
device correction of all serial numbers of the HeartMate II Left Ventricular
Assist Systems (HM II LVAS)
having Catalogue No. 1355 or 102139, which have been distributed since the
beginning of clinical studies
in November 2003. Over time, wear and fatigue of the percutaneous lead connecting
the HeartMate II
LVAS blood pump with the System Controller may result in damage that could
interrupt pump function,
require reoperation to replace the pump and potentially result in serious injury
or death. The estimated
probability of the need for pump replacement due to percutaneous lead damage
is 1.3% at 12 months,
6.5% at 24 months and 11.4% at 36 months.
Patients who are currently being supported by a HeartMate II LVAS should contact
their doctors, who can
assess the wear and fatigue of the percutaneous lead as well as provide proper
instruction on
management and care of the lead.
Thoratec is voluntarily issuing an Urgent Medical Device Correction notice
after confirming 27 reports
where wear and fatigue to the percutaneous lead necessitated pump replacement.
These reports
occurred over five years of clinical experience with 1,972 implants. All patients
who have undergone a
replacement of the HeartMate II pump survived the operation and were alive
at least 30 days
postoperatively. In five cases, pump replacement was not feasible and the patients
expired.
The affected systems were distributed to 153 hospitals and distributors throughout
the United States,
Europe, Canada and other countries. The HM II LVAS can be identified by the
catalogue number located
on the label of the package.
Hospitals are being sent an Urgent Medical Device Correction letter identifying
the probability and
symptoms of the problem, and recommending that the pump be replaced as soon
as possible if damage
to the percutaneous lead is confirmed. Hospitals are also requested to review
the instructions for care of
the percutaneous lead with their ongoing HeartMate II LVAS patients. Hospitals
with ongoing HeartMate
II LVAS patients should contact Thoratec for further instructions if they do
not receive the Medical Device
Correction letter. The labeling for the HeartMate II LVAS will be revised with
the updated risk information
related to percutaneous lead damage. Copies of this press release may be found
on Thoratec’s website,
www.thoratec.com, under Investor Relations/Press Releases.
Clinicians and patients with questions may contact the company at 1-800-528-2577,
or if calling from
outside the USA, 1-925-847-8600 (7 days a week, 8-5 Pacific Time).
FDA has been informed of this action.
Adverse reactions or quality problems experienced with the use of this product
may be reported to the
FDA’s MedWatch Adverse Event Reporting program either online, by regular
mail or by fax.
• Online: www.fda.gov/medwatch/report.htm
• Regular Mail: use postage-paid FDA form 3500 available at:
www.fda.gov/MedWatch/getforms.htm. Mail to MedWatch 5600 Fishers Lane, Rockville,
MD
20852-9787
• Fax: 1-800-FDA-0178