FDA Warns About Risk of Wearing Medicated Patches During MRIs

Subscribe to our Medical Devices Recalls RSS Feed
Date of Recall:
Thursday, March 5th, 2009

Recall Details:

Certain adhesive patches that deliver medication through the skin have been found to be a risk to patient safety. The patches, if worn while undergoing magnetic resonance imaging scans or MRIs, can cause skin burns, says the U.S. Food and Drug Administration today.

The patches of concern include both brand name and generic products and patches purchased over the counter without a prescription.

The FDA issued the Public Health Advisory on transdermal drug patches after learning that a warning was missing on some patches that contain aluminum or other metals in their non-adhesive backing. The backing is the portion of the patch not in direct contact with the skin. While not attracted to the magnetic field of the MRI, the metal can conduct electricity, generating heat which can cause burns. Users of the patches reported receiving skin burns at their patch site when wearing the patch during an MRI scan.

"The risk of using a metallic patch during an MRI has been well-established, but the FDA recently discovered that not all manufacturers include a safety warning with their patches," said Janet Woodcock, M.D., director of the FDA's Center for Drug Evaluation and Research. "Because the metal in these patches may not be visible and the product labeling may not disclose the presence of metal, patients should tell both their health care professional and their MRI facility that they wear a medicated adhesive patch."

The FDA was alerted to the missing MRI warning on Teva Pharmaceutical's fentanyl transdermal system in January. The FDA investigated and found that a similar warning was also missing on a variety of skin, or transdermal, patches delivering medications.

The FDA is reviewing the labeling and composition of all medicated patches to ensure that those made with materials containing metal provide a warning to patients undergoing an MRI and will alert the public when this information has been added. Until then, the agency recommends that people wearing medicated skin patches, including nicotine patches, talk to a health care professional about their patch at the time they receive their MRI referral. The professional will advise the patient about when to remove the patch before the procedure and about replacing it after the procedure.

Patients should also tell their MRI facility that they are using a patch when they call to schedule their appointment and should repeat this information when filling out their health history questionnaire after arriving for their appointment.

Note to TV reporters: please contact the FDA's press office if you are interested in B-roll material.

  User Comments  
There are currently no comments for this recall. Be the first to add a comment!
Click here to add a comment about this recall.
  Advertisements  
  Topical Videos  
All video content powered by YouTube.

An AP report concerning the numerous recent peanut butter recalls as well as identifying a possible source of the salmonella outbreak Read More...
  Interesting Stats  
Most Vehicle Recalls
by Make and Model
  Popular Searches  
  Related News  

Au'Some, Inc. Initiates Voluntary Recall of Certain “Transformers® Crunchy ...
Because the health and safety of children are our highest priority, Au'some is committed to taking every necessary step to ensure the safety and quality of ...
Statement of Egg Safety Center on Egg Recall by Wright County Egg
The Egg Safety Center and the Food and Drug Administration recommend that eggs should be fully cooked until both the yolks and the whites are firm, ...
Ways to Survive a Product Recall
Earlier this month, an investigation by the Food and Drug Administration revealed widespread salmonella contamination of eggs across the United States, ...