Interesting Stats | Contact Us | Free Content  
 

Abbott Issues Voluntary Recall of POWERSAIL® Coronary Dilatation Catheters

Subscribe to our Medical Devices Recalls RSS Feed
Date of Recall:
Thursday, July 30th, 2009

Recall Details:
- ABBOTT PARK, Ill. - July 28, 2009 - Abbott has conducted a voluntary recall of three lots of POWERSAILŪ Coronary Dilatation Catheters from United States distribution and one lot from international distribution as a result of four complaints (one from each lot) that the distal shaft of the catheter exhibited damage. While the issue could be detected and avoided during the preparation for use of the product, it may cause a leak of contrast material during use, which could lead to catheter functional failures and clinical consequences, including air embolism and myocardial infarction, which has the potential to lead to death.

Abbott Vascular's sales representatives contacted all customers affected by this action and instructed them to cease use of any units from the part numbers and lot numbers listed below. All outstanding units are in the company's possession or are in transit. Patients who have already been treated are not affected by this action.

Product Designation Product Number Lot
Number
Expiration Date
POWERSAILŪ 3.25x18mm (U.S.)

1005524-18

7101051

09-2009

POWERSAILŪ 4.0x8mm (CE)

1005726-08

7112051

10-2009

POWERSAILŪ 2.75x18mm (U.S.)

1005522-18

8012151

12-2009

POWERSAILŪ 3.25x8mm (U.S.)

1005524-08

8053061

04-2010

Three out of the four complaints of distal shaft damage resulted in no adverse patient effects. One complaint reported that the patient had evidence of a post-procedural myocardial infarction. Subsequently, the patient was reported to be doing well.

The United States Food and Drug Administration (FDA) has been apprised of this action, and it has classified this action as a Class I recall.

The POWERSAIL Coronary Dilatation Catheter is distributed for prescription use only to approved U.S. and international health care organizations. Affected devices can be identified by the part number and lot number combinations shown above.

Customers with questions or concerns should contact their Abbott Vascular Representative or call the company at 1-800-227-9902.

Any adverse reactions experienced with the use of this product, and/or quality problems should also be reported to the FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch website at www.fda.gov/medwatch.

  User Comments  
There are currently no comments for this recall. Be the first to add a comment!
Click here to add a comment about this recall.
  Advertisements  
  Topical Videos  
All video content powered by YouTube.

A CNN report looking at Chinese food standards and exporting of potentially dangerous food Read More...
  Interesting Stats  
Most Consumer Recalls
by Country
  Popular Searches  
  Related News  

Massive Egg Recall Raises Food Safety Questions
The outbreak may give new momentum to an effort to update US food safety laws. Ohio resident Randy Napier lost his mother in another recent food safety ...
Ground Beef Recall Affects MA BJ's Wholesale Clubs
Along with the massive nationwide egg recall due to a Salmonella outbreak, food safety experts say it's time for change. "I think we're at a point in time ...
Wary consumers, restaurants make changes after egg recall
Investigators from the US Food and Drug Administration have yet to determine the cause of the salmonella outbreaks at Wright County Egg, a farm in Iowa that ...