Pointe Scientific, Inc Issues Nationwide Recall of Liquid Glucose Hexokinase Reagent (G7517)

Date of Recall
2009-10-30

Recall Details
- October 28, 2009 - Pointe Scientific, Inc, Canton, MI is initiating a nationwide recall of all size kits of Liquid Glucose Hexokinase Reagent catalog number G7517. The reagents have been found to fail linearity at >200mg/dL that results in inaccurate glucose values above this range.

Distributors and testing laboratories who have received the Pointe Scientific, Inc Liquid Glucose Hexokinase Reagent (G7517) which is being recalled should destroy remaining inventory.