TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard
Ondal AC2000 (PDI) Television Mounting System
Date of Recall
2012-07-18CPSC Release Number
12-742Manufactured In
GermanyUnits
215 distributed to 5 hospitalsDistributor
TRUMPF Medical Systems Inc., of Charleston, SCImporter
Ondal Medical Systems of America Inc., of Sandston, VA and TRUMPF Medical Systems Inc., of Charleston, SCSold By
TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.Problem/Hazard
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.Injuries/Incidents
The firm received one report of a mounting system falling while being positioned. No injuries have been reported.Description
The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.Remedy
Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.Consumer Contact
For additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email [email protected]Recall Comments
Be the first to add a comment!